Metrology
Metrology
Millimar 1340 Inductive Probe
EN
Product features
  • Can only be used in conjunction with Millimar 1240
  • Unprecedented measuring accuracy and minimum linearity error < 0.01 %, i.e. 0.4 µm over the total measuring range
  • Probe is protected against dirt and moisture, thus ideally suited for production related application
  • Technical data

Order no.Measuring range fromMeasuring range up toDistance of lower stop MINDistance of lower stop MAXDistance of upper stop MAXLifter / retractionMeasuring force (N)Increase in measuring forceSensitivity deviationRepeatability fw (µm)Hysteresis fu (µm)Linearity deviation within +/- 1,0 mmLinearity deviation within +/- 2,0 mmIP protection categoryCable lengthWorking temperatureWorking temperature MINWorking temperature MAXOperating temperature MinOperating temperature MaxStorage temperature MINStorage temperature MAXTemperature coefficientCompatibility
5313400
-2 mm
2 mm
-2.2 mm
0 mm
3 mm
Vacuum Lifter
0,75 N
0,08 N/mm
0.3 %
0.08 µm
0.08 µm
0.15 µm
0.4 µm
IP 64
1.5 m
20 °C
10 °C
40 °C
-10 °C
80 °C
-10 °C
80 °C
0.6 µm/°C
Mahr 1340